Healthcare
ETube voltage settings |
20-35kV±2kV |
mAs settings |
Small focal spot:10-400mAs, Large focal spot:10-500mAs (Optical 600 mAs in GB) ±4mAs or ±10% (Whichever is larger) |
Focal spot dimensions |
Small focal spot:0.1mm, Large focal spot:0.3mm |
Target angle |
10° (small focal spot)/16° (large focal spot) |
Anode material |
Molybdenum (Mo) tube |
SID |
65cm (Contact imaging)/114cm (PCM imaging) |
Filter (thickness) |
30μm (Mo)/25μm (Rh) |
Maximum anode heat capacity |
300kHu (210kj) |
Rotation of swivel arm |
Contact imaging:+180° to -135° PCM imaging:+180° to -135° |
Imaging modes |
AEC equipped (handles both Contact and PCM imaging) |
AEC detector |
Sensor:48 automatically selected sensors Density adjustment:15 density steps (5% change in dose conversion per step) |
Autorotation function (recreation of identical angles for left and right MLO) |
Automatic angle setting available |
Automatic ejection of compression plate function |
Equipped |
Irradiation results obtained |
Focal spot size, tube voltage, tube current, irradiation time, mAs values, breast thickness, compression force, C arm angle, etc. |
Unit dimensions |
Main unit:W755mm×D910mm XH (maximum value) 2325mm |
Unit weight |
Main unit:200kg |
Power capacity |
208-240V~, 50 or 60Hz ±10%-Single phase |
* The above specifications may change without notice.